ANESTHESIA MACHINE & VENTILATOR MAJ – V
WITH 2 VAPORIZER GRADUATED ENFLUREN AND ISOFLUREN
MAJ – V Anesthesia Machine & Ventilator, is a new
popular type machine, has not only made a big step
in increasing its reliability, safety and combination
but also, in design, given room to equip anesthesia
ventilator and monitoring to form a complete
versatile anesthesia machine thus provides more
useful functions.
ISO standard circle system can meet the needs for
close, semi-open, semi-closed (1kpa – 4kpa) system
anesthesia. The machine has a Graduated/
ungraduated vaporizer with adjustable anesthetic
concentration. The machine is easy in operation.
• New design of flow meter with ORC (oxygen Ratio
Control) to guarantee at leas 30% O2 concentration
• Automatic N2O cut off
• ISO standard circle system fit table to the left or
the right side of the machine makes operation
more convenient in either side
• Easy observation of respiratory pressure from
pressure gauge
• With Enflurane, Isoflurane graduated vaporizer or
ungraduated Flutene , enfluren and isofluren
vaporizer
• Attachable cylinder bracket
Specifications:
1. Gas supply ( central supply) : O2 : 0.27 ~ 0.55Mpa
N2O : 0.27 ~ 0.55Mpa
Cylinder supply : O2 N2O : 0.45Mpa
(out put after decompressing)
Rated operational pressure : 0.4Mpa
2. Adjustment of flow : O2 : 0.1 ~10L/min ±10% (D1L/min ±30%)
N2O : 0.1~ 10L/min ±10% (D1L/min ±30%)
Oxygen proportion control (ORC) : O2 E25%
3. Oxygen supply : 35 ~ 75L/min
4. Circle system : close semi close (1kpa – 4kpa), semi open.
Resistance of inhalation or exhalation valves : D 0.6kPa (at 30L/m flow)
5. Alarm for oxygen deficiency:
Oxygen supply pressure lower than 0.2mPa ±10%, Alarm not more than 7 seconds, as the pressure
of oxygen is under 30 % – 50 %
6. Automatically cut – off N2O
The output of N2O, keeps decreasing till being cut off finally, as O2 is deficient
7. Concentration of vaporizer : the concentration can be adjustable, at temperature 20° oxygen supply
1L/m maximal concentration of vaporizer not more 7%
8. Ventilator at anesthesia machine : tidal volume 100ml ~ 1500ml, IPPV 6 ~ 60 x/minute, maximal 6
~ 99x/minute,I/E ratio 1 : 1 ~ 1 : 4, Inspirator flow 0.1L/s ~ 1.0L/s, peak pressure 1 ~ 6 kpa , lower
pressure 0.4 ~ 4kpa, lower this pressure 4 ~ 15s will be alarm
9. Oxygen cylinder adult 50 ~ 1500ml, child 50 ~ 300ml
10. Ventilation control : tidal volume 100 ~ 1500ml, display volume 0.05L ~ 2L, airway pressure –2 ~ 10kpa
11. volume : 44cm x (45 – 80)cm x 126cm
12. weight : 40kg

 